CTC Analytics A200S Autosampler
- A200S autosampler Technical specifications
- Sample Tray: Removable sample trays with the possibility of temperature control
- Number of samples: 200 Vials: 0.7 ml; 105 Vials: 1.5 ml (optional); 32 vials: 5 ml (optional)
- Minimum sample Vol: 0 µl for 3-fold injection
- Dosage Method: Stepper motor driven syringe of 10 ul (Hamiliton 701 type). Volume selectable in 0.1 microliter steps. Simply adjustable dead volume.
- Syringe Cleaning: Number of cleaning cycles with sample (0-99) & solvent (0-99)
- Injection Side: Installation for left or right hand side is possible
- Injection Point: Two different injection points selectable in method
- Reproductabilty: Splitless injection <1.0% absolute; On column injection < 1.0% absolute; <0.2% relative; Retention time better than 0.1%
- Splitter Valve Control: Valve activated 0-99 sec. before, and 0-99 sec after injection
Weight | 5 lbs |
---|---|
Dimensions | 8 × 5 × 5 in |
Condition |