This used Agilent 1100 Series G1315B DAD is in excellent condition. The Agilent 1100 Series G1315B Diode Array Detector is designed for the maximum optical performance. Flexible slit setting from fine bands to highest sensitivity, with settings of 1 nm the fine structure of benzene can be revealed – even at 0.7mAUFS. A broad slit of 8 or 16nm can be chosen to minimize baseline noise thus enhancing sensitivity.
System specifications include:
- Wavelength range: 190-950 nm
- Light source: Deuterium and tungsten lamps
- Detection type: 1024 – element photodiode array
- Short term noise (ASTM) single & mult-wavelength: ± 1° x 10-5 AU at 254 and 750 nm
- Drift: 2° x 10-3 AU / hr at 254 nm
- Linear absorbance range: > 2 AU (upper limit)
- Wavelength accuracy: ± 1 nm
- Wavelength bunching: 1 – 400 nm
- Diode width: < 1 nm
- Control and data evaluation: Agilent Chemstation for LC
- Slit width: 1, 2, 4, 8, 16 nm